Comparison of Glucose, AST, and Cholesterol Test Results Using Plain Vacutainer Tubes and Clot Activator Vacutainer Tubes

Kholidah Zafirah Inas Fatin, Edy Haryanto, Ayu Puspitasari

Abstract


Laboratory examinations play a crucial role in supporting disease diagnosis; therefore, laboratory test results must be accurate, valid, and reliable. However, errors may occur during various stages of the examination process, including the selection of blood collection tubes, which can influence test outcomes. In clinical chemistry testing, red-capped vacutainer tubes are commonly used, either plain tubes or tubes containing clot activators. The addition of clot activators should not affect the quality of analyte measurements. This study aimed to compare glucose, AST, and cholesterol levels in serum samples collected using plain vacutainer tubes and vacutainer tubes with clot activators. Serum samples were obtained from 16 respondents and analyzed at a primary healthcare laboratory in Surabaya. This study employed an analytic observational method with a cross-sectional and comparative study design. Data normality was assessed using the Shapiro–Wilk test, followed by the Independent t-test for normally distributed data and the Mann–Whitney test for non-normally distributed data. The results showed no significant differences (p > 0.05) in glucose, AST, and cholesterol levels between serum collected in plain vacutainer tubes and serum collected in vacutainer tubes with clot activators. Therefore, both types of vacutainer tubes can be used interchangeably for these clinical chemistry examinations.

Keywords


Vacutainer Tubes, Clot Activator, Glucose, AST, Cholesterol

Full Text:

PDF

References


Alanazi NA, Almehini AA, Yousef L, Al A, Almansour HS, Alharbi SF, et al. The Role of Laboratory Testing in Disease Diagnosis : A Comprehensive Review. Migr Lett. 2022;19:608–17.

Mrazek C, Lippi G, Keppel MH, Felder TK, Oberkofler H, Haschke-Becher E, et al. Errors within the total laboratory testing process, from test selection to medical decision-making - A review of causes, consequences, surveillance and solutions. Biochem medica. 2020 Jun;30(2):20502.

Berhan A, Sharew B, Almaw A, Abeje G, Tekalign N, Damtie S, et al. Types and frequency of preanalytical errors occurring in the hematology laboratory at Debre Tabor comprehensive specialized hospital, North Central, Ethiopia, 2025. Sci Rep. 2026 Feb;16(1).

Agung A, Denny G, Permana D, Agung A, Eka A, Luh N. Difference In Uric Acid Test Results Using Plain And. Multidiscip Indones Cent J. 2025;2(4):5004–10.

Devi A, Goswami B. A comparative study of sample collection tubes for routine biochemical parameters. Baghdad J Biochem Appl Biol Sci. 2023;4(1):27–38.

Ng WY, Yeo CP. Thrombin-accelerated quick clotting serum tubes: an evaluation with 22 common biochemical analytes. Adv Hematol. 2013;769479.

Masuki H, Isobe K, Kawabata H, Tsujino T, Yamaguchi S, Watanabe T, et al. Acute cytotoxic effects of silica microparticles used for coating of plastic blood-collection tubes on human periosteal cells. Odontology. 2020 Oct;108(4):545–52.

Bowen RAR, Remaley AT. Interferences from blood collection tube components on clinical chemistry assays. Biochem medica. 2014;24(1):31–44.

Kemenkes RI. Peraturan Menteri Kesehatan Republik Indonesia Nomor 43 Tahun 2013 Tentang Cara Penyelenggaraan Laboratorium Klinik Yang Baik. 2013.

Thomas MK, Muhammad Z, Prasanna T. Blood Glucose Monitoring. In 2023.

Lala V, Zubair M, Minter DA. Liver Function Tests. In Treasure Island (FL); 2024.

American College of Clinical Pharmacy. Reference Values for Common Laboratory Tests. Am Coll Clin Pharm. 2018;2018.

Di Lorenzo MS, Strasinger SK. Blood Collection for Healthcare Professionals: A Short Course. 2022.

Rifai N. Tietz Textbook Of Laboratory Medicine. 7th ed. Elsevier Enhanced Digital Version; 2023. 406–434 p.

Azizah FN. Pengaruh Variasi Waktu Penundaan Pemisahan Serum Terhadap Hasil Pemeriksaan Glukosa. J Muhammadiyah Med Lab Technol. 2023;6(2):184–91.

Rifai N. Tietz Fundamentals of Clinical Chemistry and Molecular Diagnostics - E-Book: Tietz Fundamentals of Clinical Chemistry and Molecular Diagnostics - E-Book. Saunders; 2018.

Kesuma S, Fera, Lamri. Perbedaan Aktivitas Enzim Aspartate Aminotransferase (AST) Pada Serum Segera Dan Ditunda Pemisahannya Selama 2 Jam. J Ilmu Kedokt dan Kesehat. 2024;11(2):301–6.

Wardani HDAK, Saktiningsih H. The Differences on Triglyceride Levels was Centrifuged Samples after Freezing for 30 Minutes and 2 Hours. J Biol Trop. 2024;24(2):916–20.

Pratiwi CD, Hariyanto H, Hermawati AH, Fajrin IN. Pengaruh Serum yang Disimpan Selama Lima Hari Suhu 2-8°C dengan Serum yang Diperiksa Langsung pada Pemeriksaan Kolesterol Total. Borneo J Med Lab Technol. 2022;4(2):292–6.

Damhuri PO, Hartuti Y, Ica M. Pengaruh Waktu Penundaan Pemeriksaan Terhadap Kadar Kolesterol. J Sains dan Teknol Lab Med. 2023;9(1):18–21.

Karakoyun I. Comparison of 30 Biochemical Analytes in 3 Different Blood Collection Tubes. 2020;18(2):61–9.

Giantini A, Indrasari ND, Suparmin S, Kasihanto P, Tansir AR. Comparison of Various Blood Vacuum Tubes Brand for Clinical Chemistry, Hormonal, Hematology, and Hemostasis Parameters. Bali Med J. 2023;12(3):2968–74.




DOI: https://doi.org/10.30602/jlk.v9i2.2083

Refbacks

  • There are currently no refbacks.


Copyright (c) 2026 Jurnal Laboratorium Khatulistiwa

Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License.



JURNAL LABORATORIUM KHATULISTIWA (JLK)

Jurnal Laboratorium Khatulistiwa (JLK) (Print ISSN: 2597-9523 | e-ISSN: 2597-9531 ) is a peer-reviewed open-access scientific journal published by the Department of Medical Laboratory Technology, Poltekkes Kemenkes Pontianak. This journal focuses on the development of medical laboratory science and technology.



Jurnal Laboratorium Khatulistiwa is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License .
CONTACT
✉ jurnallaboratoriumkhatulistiwa@poltekkes-pontianak.ac.id
📍 Poltekkes Kemenkes Pontianak, Jl. 28 Oktober Siantan Hulu, Pontianak, Kalimantan Barat, Indonesia



COLLABORATION